12 SCFH oxygen production at 25 PSI is produced by an ATF-12 oxygen concentrator module.
Details:
ATF Oxygen Modules incorporate proven pressure swing adsorption (PSA) principles into a unique patented design, which allows these systems to be compact, efficient, rugged, and lower in cost. ATF modules offer unparalleled design flexibility and enable applications where on-site/on-board oxygen generation was previously impractical.
A patented single rotary distribution valve built into the ATF module continually rotates at low speed by a small motor. The valve is maintenance-free, self-cleaning, insensitive to contamination, and invulnerable to wear. It directs the flow of compressed air to a group of four sieve beds (adsorption), while another four beds are purged into the atmosphere through the valve (desorption). The remaining four of the twelve beds are interconnected through the valve to equalize pressure as the sieve beds sequentially transition between adsorption and desorption. The small amplitutde pressure swings generated by the twelve sieve beds eliminate loud noise pulses, eliminate the need for a pressure regulator, and reduce compressor wear.
ATF-12
Input Air Requirements:
Flow
85 LPM (3 SCFM)
Pressure
25 psig (172 kPa)
Performance Data:
Product Flow
5.7 LPM (12 SCFH)
Oxygen Delivery Pressure
9 psig (62 kPa)
Product Concentration
90-95%
Product Dew Point
-100°F (-73°C)
Dimensions (WxDxH)
8 x 13 inches (21 x 32 cm)
Weight
9.8 lbs (4.5 kg)
Physical Connections
Compressed Air Inlet
1/2" ID tubing
Product Gas Outlet
1/4" ID tubing
Ambient Operating Conditions
Locate the oxygen concentrator system in a well-ventilated area that is protected from weather elements.
Inside room or enclosure: Temps should remain between 40°F (4°C) and 130°F (54°C)
In storage: Temps should remain between 0°F (-18°C) and 140°F (60°C)
Control Power Requirements (Single Phase)
120 V / 208-240 V ~ ± 10%, 50 / 60 Hz, 0.025 A or 208-240 V ~ ± 10%, 50 Hz, 0.025A
Typical Power Consumption
3 W
Gear Motor
7398 (FMR 1268)
Compressed Air Input Requirements
Clean Air (oil-less). Max Temp of 170°F (77°C). Performance Rated at 120°F (49°C).
Mounting
Two Integral Mounting Pads Provided.
Toe Clamps (Optional)
Vertical Mounting Preferred.
Must Be Isolated From Compressor Vibration.